Daily maintenance and cleaning maintenance of high and low temperature humid heat alternating test chamber

Qin Zhuo test equipment introduces you to the maintenance and cleaning of the high and low temperature humid heat alternating test chamber in daily work

1. How to clean and maintain the inside and outside of the high and low temperature humid heat alternating test box:

1. Before and after the operation of the high and low temperature humid heat alternating test chamber, the internal impurities should be removed.

2. The power distribution room should be cleaned at least once a year. When cleaning, you can use a vacuum cleaner to remove the indoor dust.

3. The outside of the cabinet must be cleaned more than once a year, just wipe it.

2. Inspection and maintenance of humidifier in high and low temperature humid heat alternating test chamber:

The water storage in the humidifier should be replaced once a month to ensure that the water quality is clean, and the humidified water tray should be cleaned once a month to ensure smooth water flow.

3. Check the over-temperature protector in the high and low temperature humid heat alternating test box: When the high and low temperature humid heat alternating test box is in operation, set the maximum value of the over temperature protection plus 20 ℃ ~ 30 ℃. When the temperature in the test chamber rises to the over-temperature protection set point, the heater power supply stops, the over-temperature warning light is on but the fan is still running, if it is running for a long time and is unattended, please make sure to check the over-temperature protector Set up properly.

4. Removal of the dust in the condenser of the high and low temperature humid heat alternating test chamber: The condenser should be maintained regularly on a monthly basis. Use a vacuum cleaner to remove the dust adhering to the condenser heat sink or use high-pressure air to remove the dust.

V. Replacement of wet-bulb test cloth in high and low temperature wet heat alternating test box: When the surface of the test cloth is not clean or hard, or after the temperature control is completed, the test cloth must be replaced before continuing the temperature and humidity sphericity control. The test cloth should be replaced every three months. The temperature measurement body should be wiped with a cleaning cloth when replacing. The new test cloth should be cleaned first.

6. Inspection and adjustment of the wet bulb water level of the high and low temperature humid heat alternating test chamber: The water level of the accumulator should not be too high, so that the water overflows the accumulator or is too low to make the wet ball test cloth absorb water abnormally, affecting the accuracy of the wet ball Just keep six points full. The water level of the water tube can be adjusted to adjust the height of the water box. In summary, only correct cleaning and maintenance can make the high and low temperature humidity heat test chamber exert its best performance and maximum function.